Skip to Content

BPBIO320 & BPBIO750 PumpValve Cable Type Change for Cable Durability

August 17, 2026 by
BPBIO320 & BPBIO750 PumpValve Cable Type Change for Cable Durability
Hanul Shin

We hope everything is going well with your business.


Purpose of noticeBPBIO320 & BPBIO750 Pump Valve Cable Type Change to Improve Resistance Against Pump/Valve Cable Breakage


Details of Change- Linear Sol Valve Cable Type Modification (from sold ered type → to cable type)

P8H246002 : [Assy]PumpValve BPBIO750 Renewal
BeforeAfter



Notes

No.Summary of ChangesChanges vs. Existing Materials
1InterchangeableO- Same shape and function before and after the change- Compatibility remains identical- No material distinction required
2Performance/Function ChangeX
3Distinguish from Previous MaterialX

1) This valve component is applicable only to the BPBIO320 & 750 Renewal Series.2) For CS parts shipment, the change will be applied after the existing stock is exhausted.


Implementation Date (Production)

Product NameApplication DateApplied S/NPart Subject to Change
BPBIO320 Renewal Series2025-01-14B825000132~[Assy]PumpValve BPBIO750 Renewal
BPBIO750 Renewal Series2025-01-20H225000047~