Digital Program 0424 :
InBody 970 : Click here to download
InBody BWA2.0 : Click here to download
The update improves the problem of Contact Customer Support message displaying during the measurement. We highly recommend updating devices before shipping to customers.
Change detail :
<Important>
1. BWA2.0 / InBody970 Improved the problem of Contact Customer Support message displaying during the measurement
- The cause of the issue is a program bug. When the contact of the electrodes are not good, the Analog Board fails to respond to the electrode contact status check requested by the DNO Board within 40 seconds.
The response time has been increased from 40 seconds to 120 seconds.
2. Improved 5G Wi-Fi Dongle
- Wi-Fi driver updated to prevent the device stopping during the measurement.
- Improved the problem of not connecting to the routers which are hiding SSID
<Requested by InBody USA>
1. Phantom Printer Added
- To connect to Phantom printer, the OS Program must be K003
<Requested by InBody Japan>
1. Changed Phrase of SMI
- Phrase changed on the right side of the LCD as well as the result sheet.
<Requested by InBody Mexico>
1. InBody Mexico service center details displayed in Spanish-Mexico is selected.
2. Fixed voice clipping issue in Spanish-Mexico
<Requested by InBody Japan>
1. LCD/Result Sheet image changed in Chinses
<Common Application>
1. Countries added
- Cyprus, Croatia, Slovenia, Tunisia, Albania, Macedonia, Kenya, Malta, Moldova, Pakistan, Palestine, Jordan, Libya, Oman, Bahrain, Kuwait, Qatar, Iraq, Yemen, Philippines, Brunei, Myanmar, Mongolia
2. Phrase in the UI changed when Italian is selected
3. Manual Input of the Weight Circumference added
- After measuring the waist circumference, a confirmation screen is added and can be modified.
- When the Done button in pressed on the Y-Scope without measuring, direct input screen is displayed.
4. LCD/Result Sheet image changed in Portuguese-Brazil
Please also note the following changes from 0423
1. (request 22-0433) Phantom printer added (US corporation requested)
- Added phantom printer to printable printer list.
4. (Request 22-0278) Load cell calibration result pass/fail criteria added (requested by the Ministry of Production)
- When performing load cell calibration during production without 0kg, the initial value is recorded as -nan. At this time, it is modified to be displayed in red on the QC Report screen.
6. (Request 22-0457) Added a function that does not reset country/language settings when setup is initialized at the request of the production department
- Setup2 initialization is added under the existing Setup initialization after entering initialize in Test Mode
7. (request 22-0457) Fixed the problem of audio being cut off on the posture screen in Spain and Mexico
8. (Request in progress) Addition of countries
[OS Program K003Ver]
InBody970/InBodyBWA2.0 : Click here to download
<Requested by InBody USA>
1. Phantom Printer Driver Added
- K002 version can be used if neither Phantom nor Brother Printer is used
Date of Implementation