Dear InBody Team,
1. Change Classification (Reason)
- Quality Improvement
2. Changed Details- 24/8/13~8/20, in the 380DM-0421 version program applied to production, there was an issue wherethe results sheet was labeled as 'InBody580.' This issue has now been corrected.※During that period, 124EA of the US product were produced.
- During the production period from 24/8/21 to 8/28, the program version 380DM-0422ver was applied.An issue where the keypad became unresponsive or had a delayed touch response in NAWI mode orwhen set to Japanese was identified and has been resolved※There was no production of NAWI or Japanese market products during that period
3. Date of Implementation
| Model | Program | S/N | Manufacture Date |
| InBody380 | Digital Program(0423Ver) | P924002538 | 29.08.2024 |
4. CS Plan
- InBody380 products produced from 24/8/13 to 8/20 (Digital Program 0421ver)
must be upgraded to Digital Program 0423ver before shipment.
The full details are available in the attached Design Change document.
Design Change Letter IB-24-25 Change in InBody380 Digital Program (0423Ver).pdf