Change in the Digital program - 270/470DM-0423Ver - InBody270 OLD LCD

17 sierpnia 2026 przez
Change in the Digital program - 270/470DM-0423Ver - InBody270 OLD LCD
Hanul Shin

Note : This program allows to use INBT500 with the OLD LCD InBody270


The OLD LCD version of the InBody270 received a new update which contains the same function as the NEW LCD InBody 270 (0424 version). 


InBody270/470 (270/470DM 0423Ver) can be downloaded here


Change details :


[0385 Digital Program Correction]

  1. Due to the problem of Russian/Thai voice being cut off on the posture screen, the part where the total language measurement time was increased by 5 seconds was re-corrected.
  2. Due to the addition of the open protocol retransmission function in version 0383, a part to save retransmission data was added, and a problem of delay of about 4 seconds was found in that part, and the part that was delayed about 4 seconds was corrected.
  3. Modifications related to QR code recognition function  - Fixed a problem that was not recognized if there was a : in the middle of the data when using base64 encryption."

"[0390 Digital Program Correction]

[UI change]

  1. Added a new country (request 20-0718)
  2. - Algeria, Armenia, Austria, Azerbaijan, Belarus, Belgium, Estonia, Georgia, Iceland, Kazakhstan, Latvia, Lithuania, Luxembourg, Morocco, British Virgin Islands, US Virgin Islands
  3. Unit change
  4. -lbs -> lbs 
  5. -ft. in. -> ft in
  6. Added unlock function with barcode. (request 20-0511)
  7. - By adding a barcode unlock function to the screen lock function, when the screen is locked after a set time, the lock is released when an ID is entered with a barcode, and a function that can be measured only with the corresponding ID is added.
  8. Modification of serial number input function.
  9. - Modified to input product S/N with GS1-128 barcode when inputting with barcode.
  10. - Modified to allow input of up to 10 digits, which was previously fixed to 9 digits.
  11. Modified to show the number of resources in the QC report.
  12. In case of NAWI weighing equipment, modification to prevent use of weight input mode (request 20-0692)
  13. When logging in to LB WEB, the password is displayed as *, and when modified, the existing data is deleted (request 20-0630)
  14. Modifications related to wireless printers
  15. - In addition to Samsung wireless printers that could be interlocked previously, other undeclared wireless printers were modified to be searched as ""Unknown Printer"".
  16. - Modified the part that took about 35 seconds to print to about 15 seconds.
  17. InBody X60 equipment open protocol data restriction request (request 20-0662)     - Modified to output as default item value when outputting X60 open protocol
  18.  Added function to receive two blood pressure data (request 20-0503)
  19. - Added function to receive up to two blood pressure data and output them as screen display and result sheet/CSV file

[clinical]

  1. Modifications related to result sheets and CSV output items (request 20-0134)
  2. - Added lean mass, SMM/WT, and recommended calorie intake (direct input) items to the right side of the result sheet, thermal result sheet, and CSV file.
  3. - Changed the research item size to [2] and grouped it into research items like Korean.
  4. Fix sentence errors
  5. - Application of Thai language modification requested (request 20-0548)
  6. - Application of Portuguese language modification request (request 20-0641)
  7. - Application of the part requested to be modified in French (request 20-0666)
  8. - The resource was pre-deployed in an emergency.
  9. Modification of body shape evaluation sentence (request 20-0505)
  10. - Apply the language correction for the country as follows
  11. - Above BMI standard / PBF standard: Plump --> Overweight
  12. - BMI standard / PBF standard or higher: plump --> mildly obese"

[0391 Digital Program Correction]

  1. Fixed a problem that only 9 digits were recognized when recognizing the serial number as a GS1-128 type barcode for the US in version 0390.     
  2. - (21) Modified so that if the serial number is 9 digits, 9 are read, and if 10 digits, 10 digits are entered.
  3. Fixed a typo on the warm-up screen     
  4. - Modify empitical -> empirical in the bottom line""[0400 Digital Program Correction]

[UI change]

  1. Add printer language  
  2. - Added PCLmS printer language.
  3. Domestic open protocol function modification  
  4. - Fixed a problem that Open Protocol was disabled when Save was pressed in Preferences No. 1 after setting Open Protocol in Korean.
  5. DC mode added (request 20-0782)  
  6. - Function to transmit the DB measured by LB120.  
  7. - If you enter environment setting No. 7 DB copy and press keypad '0' 5 times on the copy all/select copy selection screen, (DC MODE) is displayed on the screen and DC mode is set.  
  8. - When copying is executed, the DB measured by the interlocked LB is transmitted one by one.
  9. Changed the function that received blood pressure data twice to the hidden function.  
  10. - When the blood pressure monitor is linked, an image of """"The blood pressure monitor is connected"""" is displayed on the home screen. If you press the upper left corner of the image 5 times, the function of receiving two blood pressure data is activated.  
  11. - By default, it operates as a function that receives one blood pressure data, measures it, and prints it on the result sheet.
  12. LAN Mac address change function added (request 21-0018)  
  13. - Enter environment setting in TEST MODE, connect LAN at """"Internet Connection No. 16"""", and press No. 5 on the top left to display a screen for changing LAN Mac Address.  - If necessary, change one digit at the end and save it, the device will be rebooted and the changed MAC Address will be applied. 
  14. - If you are in an environment where you cannot connect to the Internet, change the MAC Address after setting and saving an arbitrary IP in the mode of directly entering the IP.
  15. FFMI, FMI CSV file location change (requested by Japanese corporation)  
  16. - FFMI and FMI, which were not previously output to the CSV file, were added in the middle in the 0390 version, but they were removed again and the location was changed to the back.
  17. Other fixes  
  18. - Add ID to transmission data file name when open protocol image data transmission function (requested by Japanese corporation)  
  19. - Modified to display ""---" in case of edema when InBody score is output on the thermal result sheet (requested by Japanese corporation)

[clinical]

  1. Change the BMI evaluation item check criteria when BMI criteria is set to Asian (request 21-0089)  
  2. - When BMI standard is set to Asian    Before Asian change: underweight less than 18.5, normal 18.5~23, overweight 23~30, severe overweight over 30    After Asian change: underweight less than 18.5, 18.5~23 normal, 23~25 on task, over 25 severely overweight    changed to."

[0405 Digital Program Correction]

[UI change]

  1. Scenario modification when interlocking height gauge with LB. (request 21-0323)  - When receiving the height value from LB, the height value of InBody is modified to be initialized.
  2. Europe/USA/China privacy agreement update (request 21-0343)  
  3. - For Europe (Dutch, Swedish, Spanish, English, Italian, French, Finnish, German)/China, the contents of the privacy agreement have been updated.  
  4. - In the case of the United States (Canada), the contents of the Privacy Agreement and Terms of Use Agreement have been updated.
  5. Open protocol existing bug fix  
  6. - Fixed the problem that the DB cannot be saved after measuring about 800 times when setting the open protocol bi-directionally.  
  7. - Fixed an issue where retransmitted data was stored incorrectly when measuring after setting open protocol bi-directional.  
  8. - After backing up the existing equipment DB, upgrade to the corresponding program, and restore the DB while open protocol bi-directional is set, the retransmission data that was stored incorrectly is initialized.
  9. Added a function that pops up when saving DB fails. (Request 21-0080)  
  10. - After saving the DB, it compares with the DB just saved and checks whether it is saved correctly.  
  11. - If saving fails, an error pop-up window is displayed to contact the customer center.
  12. Update to QR code rev.3  
  13. - Modified the number of digits related to body fat percentage by body part from 4 digits to 5 digits so that it does not break even in the case of severe obesity.
  14. Apply the request of the Ministry of Production (request 21-0342, request 21-0136, request 21-0164)  
  15. - Modified so that Auto Zig Test and Kim Gu-hyun Test (Aging Test) can be performed at once after Auto Zig integration. (Zig Version: KTM.00.00.09)  
  16. - Modified to show reactance and phase angle on History result sheet when QC report is output (770).  
  17. - Modified so that the currently set country/language appears in the QC report.  
  18. - Modified to show the number of resources currently held in the QC report.  
  19. - If the keypad test does not pass, it is modified so that it does not pass even if all of the IF tests pass.  
  20. - If the load cell is not calibrated, it is modified to be displayed in red.
  21. Italian correction (request 21-0331)  
  22. - Change from Servisio to Servizio on the LCD cloud connection screen
  23. Added memory space check function  
  24. - When you touch the version 5 times on the first page of the environment setting, the detailed version is displayed at the bottom. At this time, the entire memory space and the space in use are displayed.
  25. Fixed the problem that analog board version name disappears when entering test mode.  
  26. - Fixed the problem that only the analog board version name disappears when entering the test mode in the 0400 version."

[0416 Digital Program Correction]

  1. Added other setting function to the 4th page of environment setting  
  2. - (Request 21-0417) Addition of on/off setting function for redisplay/reprint button after measurement  
  3. - (Request 22-0018) Added all/recent setting function in the history part of the result screen after measurement
  4. (Request 21-0607) When rebooting after logging in to LB WEB, press X on the re-login screen to change settings, modifying to enter the password.
  5. (Request 21-0662) Age limit changed from 99 to 150.
  6. (Request 22-0017) Correction of misspelled country/language names.      
  7. (country part)      Morocco -> Morocco      Belgium -> Belgium      
  8. (language part)      Netherlands -> Dutch
  9. (Request 22-0151) Add Brother printer to DNO2.      
  10. Added as a language called Brother GDI when setting up the printer.
  11. (Request 21-0564) Changed to save and load from external USB when DNO backup/restore in test mode.
  12. (Request 22-0150) QC data server transmission function​



Change in the Digital program - 0424 - BWA2.0/970