Przejdź do zawartości

[Design change] Change in InBodyJ30/J50 Motor Program

17 sierpnia 2026 przez
[Design change] Change in InBodyJ30/J50 Motor Program
Hanul Shin

Dear InBody Team,

1. Change Classification (Reason)

- Quality Improvement


2. Changed Details- The motor program has been changed to address the issue that the previous measurement value wasdisplayed when height measurement was completed using the pressure mode in the J30M-113ver. It hasbeen corrected to ensure normal measurement even when using the pressure mode.


4. Date of Implementation

ModelProgramS/NApplication Date
InBodyJ30J30M-114verC5240001102024-08-27
InBodyJ50J30M-114verC9240000712024-08-27


The full details are available in the attached Design Change document.

Design Change Letter IB-24-26 Change in InBodyJ30,J50 Motor Program (114Ver).pdf